We provide professional optical product testing services (laser product safety testing and performance parameter testing, photobiological safety testing for non‑laser products), as well as energy efficiency testing services (EU ErP energy efficiency testing, US DOE energy efficiency testing, California CEC energy efficiency testing), among others.
If you have requirements for laser products (laser product safety testing and laser performance parameter testing) and energy efficiency testing (EU ErP energy efficiency testing, US DOE energy efficiency, California CEC energy efficiency, UK energy efficiency testing), please feel free to email us at: admin@ctnt-cert.com
When laser hair removal devices enter international markets (especially the EU CE certification system), the IEC 60601‑2‑22 test report is the core technical document proving compliance with the dedicated laser safety requirements for medical electrical equipment.
The fourth edition of this standard (IEC 60601‑2‑22:2019) has explicitly included aesthetic laser equipment within its scope and added coverage requirements for Class 1C laser devices. For laser devices with hair removal as their intended use, obtaining this report is not only a necessary step for compliance and market access but also third‑party endorsement of product safety and effectiveness.
The operating principle of laser hair removal devices determines their risk profile: they act directly on human skin and involve high‑energy laser radiation. The international regulatory framework classifies them as medical electrical equipment or electrical products with medical applications, requiring compliance with dual standards – the general safety standard IEC 60601‑1 and the laser equipment‑specific standard IEC 60601‑2‑22.
IEC 60601‑2‑22 is the "dedicated examination paper" for the electrical safety and laser radiation safety of laser hair removal devices. The test report is a foundational technical document for CE certification, FDA registration, and NMPA registration.
1.1 What the Standard Is
The full title of IEC 60601‑2‑22 is "Medical electrical equipment – Part 2‑22: Particular requirements for the basic safety and essential performance of laser equipment for surgical, aesthetic, therapeutic and diagnostic use". It supplements the general standard IEC 60601‑1, setting additional requirements specifically for the special risks of laser medical electrical equipment.
1.2 Why Laser Hair Removal Devices Fall Within the Standard's Scope
The standard explicitly applies to "laser equipment for aesthetic or veterinary applications", and hair removal is a typical aesthetic application scenario.
The laser product categories covered by the standard are Class 3B or Class 4, as well as Class 1C laser products where the enclosed laser is Class 3B or 4. High‑energy laser hair removal devices typically fall within this range.
It should be distinguished that low‑energy home‑use IPL (intense pulsed light) devices may fall under IEC 60335‑2‑113 (household appliance safety standard), while laser hair removal devices, due to their laser product classification, fall under IEC 60601‑2‑22.
1.3 Key Updates in the Fourth Edition
Important changes in the fourth edition compared to the third include: inclusion of Class 1C laser equipment, exclusion of LED products (now covered by IEC 60601‑2‑57), and alignment with IEC 60601‑1:2005/AMD1:2012 and IEC 60825‑1:2014. When applying for a test report, confirm that the testing body is using the latest version.
2.1 Laser Radiation Safety
Laser product classification: Equipment must be correctly classified in accordance with IEC 60825‑1; hair removal devices are typically Class 3B or 4.
Protective housing and interlocks: The beam path housing must be equipped with interlock devices; once opened, laser emission must terminate immediately.
Key switch: Class 3B/4 equipment must be activated via key or password to prevent unauthorised use.
Emergency stop device: Must be equipped with an easily accessible emergency stop function, with response time typically required to be ≤1 second.
Beam delivery system: For fibre‑optic delivery systems, connection reliability must be ensured, with safety measures in the event of breakage.
2.2 Output Control and Parameter Accuracy
Laser output parameters (wavelength, power, energy density, pulse duration) must be consistent with nominal values, with errors typically controlled within ±10% to ±20%.
The laser "ready" indicator (visual and audible) and "emitting" indicator must be clear and unambiguous.
Output parameter settings and displays must be precise to ensure controllability of treatment dosage.
2.3 Electrical Safety
Basic electrical safety items such as leakage current, earthing resistance (≤0.1 Ω), and dielectric strength testing (e.g., 1500 V/60 s without breakdown) follow IEC 60601‑1.
Electromagnetic compatibility items such as electrical fast transient/burst immunity are also within the testing scope.
2.4 Biocompatibility of Applied Parts
Accessories such as handpieces and probes that come into direct contact with patient skin must be assessed for cytotoxicity, skin irritation, and sensitisation in accordance with the ISO 10993 series.
3.1 Self‑Assessment Before Application
Confirm the product laser classification (Class 3B/4 or Class 1C).
Confirm the applicable standard version: IEC 60601‑2‑22:2019 or EN IEC 60601‑2‑22:2020 (EU harmonised standard).
Prepare technical documentation: product specifications, laser parameter descriptions, risk analysis report (ISO 14971), user manual, etc.
3.2 Choosing a Testing Body
The testing body must have testing qualifications for IEC 60601‑2‑22, typically determined by its CNAS accreditation scope.
Prioritise bodies that also have IEC 60601‑1 and IEC 60825‑1 testing capabilities to avoid the coordination costs of splitting submissions.
3.3 Sample Submission and Testing
The customer submits product information (equipment type, laser parameters, intended use); the testing body evaluates and issues a plan and quotation.
Sign the service agreement and provide test samples (complete device or key components).
The laboratory conducts comprehensive testing in accordance with the standard, including laser radiation safety, electrical safety, and output parameters.
3.4 Report Compilation and Timeline
After testing is completed, engineers compile the data and prepare the report, which is delivered after internal review.
The standard cycle is typically 7 to 15 working days; if corrective actions or supplementary testing are involved, it will be extended accordingly. Some bodies have a testing cycle of 2‑3 weeks, which can be shortened with expedited services.
4.1 Report Validity and Version Issues
The fourth edition IEC 60601‑2‑22:2019 has replaced the third edition (2007 version). When applying, ensure the testing basis is the currently effective version to avoid the report not being recognised by the target market due to an outdated standard.
4.2 Interface Between Standards
IEC 60601‑2‑22 cannot be used alone. It builds upon IEC 60601‑1 (general safety) and IEC 60825‑1 (laser product safety). Test reports typically need to demonstrate comprehensive compliance with all three standards.
4.3 Difference Between Home‑Use Products and Professional Equipment
If a laser hair removal device is designed as a home‑use product and meets the Class 1C definition of IEC 60825‑1, it may fall under IEC 60335‑2‑113 rather than IEC 60601‑2‑22. Manufacturers must make an accurate determination based on intended use scenario and laser classification.
The essence of the IEC 60601‑2‑22 test report is to prove to regulators and the market, using an internationally recognised technical language, that: this laser hair removal device's laser radiation is controlled, its electrical system is safe, and its parameters are accurate. For companies aiming to enter the EU or other markets adopting IEC standards, this report is not optional – it is the starting point of technical compliance. It is recommended to incorporate standard requirements at the product development stage rather than discovering design‑level non‑conformities only at the time of submission, in order to shorten the overall time to market.
Shenzhen Zhongwei Inspection is a professional laser product testing and certification body, providing IEC 60601‑2‑22 testing services for medical laser equipment. With ten years of laser equipment testing experience and over 10,000 successful cases, it offers one‑stop support for the compliant export of your medical laser equipment.